DO SENSOR MODEL OXYGOLD 120 /سنسور اکسیگلد 120
مشخصات مدل G:
مشخصات فنی مدل B:
مشخصات مدل G:
a-length | 120 mm |
Measurement Principle | electrochemical reduction of oxygen |
Measuring Range | 1 ppb to 40 ppm (DO) or 0.02 to 1000 mbar (pO₂) |
Membrane / Cap | Delivered with Optiflow |
Electrolyte | Oxylyte G |
Oxygen Consumption | Ca. 100 ng/h in air at 25 °C |
Required Flow | ≥ 0.1 m/s |
Response Time t90% | < 60 s at 25 °C |
Temperature Sensor | NTC 22 kOhm |
Current in Air at 25 °C | 180 - 500 nA |
Drift at Room Temperature | < 1 % per week |
ATEX Approval | CE 0035 II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb, CE 0035 II 1/2 D ia IIIC T x °C Da/Db |
IECEx Approval | CE 0035 II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb, CE 0035 II 1/2 D ia IIIC T x °C Da/Db |
Autoclavable | Yes |
CIP | Yes |
Steam Sterilizable | Yes |
Operating Temperature Range | 0 to 130 °C |
Pressure Range (bar g) | 0 to 12 bar |
Max. CO₂ Partial Pressure | 0.01 bar |
O-ring Material | EPDM (FDA approved, USP class VI) |
Surface Quality of Steel | Ra < 0.4 µm (N5) |
Certificate | material and performance test certificate |
Diameter | 12 mm |
Process Connection | PG13,5 |
Electrical Connector | VP6 |
Wetted Parts | Stainless steel 1.4435, Silicone, EPDM |
Electrode system | Silver platinum combination |
Polarization Voltage and Time | -670 ± 50 mV, ≥ 2 hours |
Serial Number | Yes |
a-length | 120 mm |
Measurement Principle | electrochemical reduction of oxygen |
Measuring Range | 8 ppb to 40 ppm (DO) or 0.2 to 1000 mbar (pO₂) |
Membrane / Cap | Delivered with Optiflow |
Electrolyte | Oxylyte B |
Oxygen Consumption | Ca. 100 ng/h in air at 25 °C |
Required Flow | ≥ 0.1 m/s |
Response Time t90% | < 60 s at 25 °C |
Temperature Sensor | NTC 22 kOhm |
Current in Air at 25 °C | 180 - 500 nA |
Drift at Room Temperature | < 1 % per week |
ATEX Approval | CE 0035 II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb, CE 0035 II 1/2 D ia IIIC T x °C Da/Db |
IECEx Approval | CE 0035 II 1/2 G Ex ia IIC T4/T5/T6 Ga/Gb, CE 0035 II 1/2 D ia IIIC T x °C Da/Db |
Autoclavable | No |
CIP | Yes |
Steam Sterilizable | No |
Operating Temperature Range | 0 to 100 °C |
Pressure Range (bar g) | 0 to 12 bar |
Max. CO₂ Partial Pressure | 12 bar |
O-ring Material | EPDM (FDA approved, USP class VI) |
Surface Quality of Steel | Ra < 0.4 µm (N5) |
Certificate | material and performance test certificate |
Diameter | 12 mm |
Process Connection | PG13,5 |
Electrical Connector | VP6 |
Wetted Parts | Stainless steel 1.4435, Silicone, EPDM |
Electrode system | Silver platinum combination |
Polarization Voltage and Time | 0 ± 50 mV, ≥ 24 hours |
Serial Number | Yes |